INTRODUCTION:

A precise knowledge of various magnetic parameters of ferromagnetic substances, viz. coercively, retentively, saturation magnetization and hysteresis loss, and ability to determine them accurately are important aspects of magnetic studies. 

The information about the aforementioned properties can be obtained from a magnetic hysteresis loop which can be traced by a number of methods in addition to the slow and laborious ballistic galvanometer method. Among the typical representatives of AC hysteresis loop tracers, some require the ring form of samples while others can be used with thin films, wires or even rock samples. Ring form samples are not always practically convenient to make while in others demagnetization effects sometime become quite important. 

The present set-up can accept the samples of thin wires of different diameters. The demagnetization effects, different diameters of samples and eddy currents (due to the conducting property of the material) has been taken into account within the design or graphically. 

FEATURES   : 

  • Measures magnetic parameters accurately
  • Demagnetization, eddy currents and sample cross-sectional area have been accounted for
  • Capable of detecting the number of magnetic phase present in a sample
 
Design Principle
When a cylindrical sample is placed coaxially in a periodically varying magnetic field the magnetization in the sample also undergoes periodic variation. This variation is picked up by a coil, placed coaxially with the sample. 

For the uniform field Ha produced, the effective field H acting in the cylindrical sample will be

                                                                     H = Ha–NM

                                                                          or

                                                                                                                        (1)

where M is magnetization, N is normalized demagnetization factor including 4p and J is the magnetic polarization defined by

B = μ0H+J

with B=μH or μ0(H+M) as magnetic induction. The signal corresponding to the applied field, Ha can be written as

                                                                        e1 = C1Ha                                              (2)

where C1 is a constant

Further the flux linking with the pick-up coil of area Ac due to sample of area As will be 

φ = μ0(Ac–As)H’+AsB

where H’ is the field in the excess area of the pick-up coil.
Under certain conditions this equation reduces to

φ = μ0AcH+AsJ

The signal induced in the pick-up coil (e2) will be proportional to d φ/dt  which after integration yields 

e3 = C3f = C3µ0AcH+C3AsJ)

Solving (1), (2) and (3) gives

Based on these equations the electronic circuit has been designed to give values of J and H and hence the hysteresis loop. Further different magnetic phases present in the sample may also be identified by electronically manipulating the pick-up signal.

Description of the set-up

 

The block diagram of the set-up is given above and a brief description now follows:

Basic Circuit

The magnetic field has been obtained with an ac mains driven multilayered solenoid. This magnetic field has been calibrated with a Hall Probe for uniformity and correspondence with the magnetic field calculated through ac current passing in the solenoid. A small resistance in series with the solenoid serves the purpose of taking a signal e1 corresponding to H. 

The signal e2 (corresponding to dJ/dt) is taken from the pick-up coil placed at the centre of the solenoid and contains the sample. It is integrated and corrected for phase. This signal is then subtracted from the reference signal e1 and amplified to give the signal corresponding to J. The e1 signal is also subtracted from 3e3 in correct ratio (to account for demagnetization and area ratio) and amplified to give signal corresponding to H.

e2 is also passed through the differentiator for getting signal corresponding to d2J/dt2 which is used for phase identification.

Applications:

The following magnetic parameters can be measured by this set-up:

  • Coercively
  • Retentively
  • Saturation magnetization
  • Various magnetic phase identification
  • Hysteresis loss 

The equipment is complete in all respect, including a set of samples (wires of Nickel, and different grades of iron etc.). A Cathode Ray Oscilloscope will however be required.

 

 Note: There may be any change in specification due to continuous R & D without notice.

 